Dwyer Series DMF Digital Mass Flow Controller

Programmable, RS485 Interface, up to 500 psig, ±1% Accuracy
 

Program, record, analyze, and control flow rates of various gases with Series DMF Digital Mass Flow Controller. Controllers are designed with straight tube sensors with restrictor flow elements to ensure laminar gas flow for accurate and repeatable results. Series DMF are unaffected by temperature and pressure variations.

Controllers can be programmed for various control functions including flow set point, totalizer, stop totalizer, totalize from preset flow, stop and preset total, auto zero, and more. Auto zero feature utilizes an automatic balancing circuit that virtually eliminates drift of zero and span. In addition, Series DMF controllers have two dry contact closures to operate pumps, valves, or other process equipment.

The Series DMF features an auto tune function to optimize control response for a specific gas under actual process conditions. Self diagnostic tests are run at power up to ensure the controller is in optimum working condition. Calibration information for up to 10 gases and conversion factors for up to 256 different gases are stored in memory. High and low gas flow alarm limits are programmed via RS-485 interface. Software supports programmable flow modes, allowing execution of custom programming of up to ten steps.  Dimensional Information

 
SPECIFICATIONS
Service: Clean gases compatible with wetted parts.
Wetted Materials: 316 stainless steel, 416 stainless steel, Viton® O-rings.
Accuracy: ±1% FS including linearity between 59 to 77°F (15 to 25°C) and 10 to 60 psia (0.7 to 4 bar); ±2% FS from 32 to 122°F (0 to 50°C) and 5 to 150 psia (0.3 to 10 bar).
Repeatability: ±0.15% of full scale.
Response Time: 0.6 to 1.0 second to within ±2% of setpoint over 20% to 100% FS.
Output: Linear 0-5 VDC (2000 Ω min. load impedance); 0-10 VDC (4000 Ω min. load impedance).
Maximum Particulate Size: 100 microns.
Temperature Limits: 41 to 122°F (5 to 50°C).
Power Supply: ±15 VDC; 13.5 watts maximum.
Process Connections: 1/4" compression fitting for flow rates ≤50 L/m; 3/8" for flows ≥60 L/m.
Pressure Limits: 500 psig (34.5 bar).
Leak Integrity: 1 x 10-9 sccs of Helium.
Computer Interface: RS-485.
Agency Approvals: CE.
 
ACCESSORIES
DMF-1100P -110V Power Supply with 25-pin connector
DFM-220PE - 220V Power Supply with 25-pin connector
DMF-CBL1 - 6 ft cable with branch to separate power supply
DMF-CBL2 - 6 ft cable with branch to computer port
 
POPULAR MODELS:
Model Number Range¹
 DMF-41401 0-10 sccm
 DMF-41402 0-20 sccm
 DMF-41403 0-50 sccm
 DMF-41404 0-100 sccm
 DMF-41405 0-200 sccm
 DMF-41406 0-500 sccm
 DMF-41407 0-1 l/min
 DMF-41408 0-2 l/min
 DMF-41409 0-5 l/min
 DMF-41411 0-15 l/min
 DMF-41431 0-30 l/min
 DMF-41433 0-50 l/min
 DMF-41842 0-100 l/min
 