Model
660 Air Velocity Monitors are a practical, affordable way
to continuously monitor for safe airflows through laboratory
fume hoods. They are typically installed in the fume hood side
fascia and connected to the interior sidewall via 1-1/8" flexible
tubing. As the exhaust fan draws air through the device, a
sensitive constant temperature thermistor measures flow and
lights a green (normal), yellow (high) or red (low) LED. An
audible alarm also warns of low flow and requires manual resetting.
Mounting holes fit standard single gang electrical box. Dimensional
Information
SPECIFICATIONS:
Service: Air and non-combustible,
non-corrosive gases. High Setpoint Range: 0-275 FPM (0-1.397 m/s). Low Setpoint Range: 0-150 FPM (0-.792 m/s). Repeatability: ±7% of full span, 0-50 and 150-275 FPM; ±5%
of full span, 50-150 FPM. Compensated Temperature Range: 50 to 90°F (10 to 32.2°C). Operating Temperature Range: 32 to 120°F (0 to 48.9°C). Power Supply: 24 VAC, 3 watts maximum. Response Time: 6-10 seconds. Warm-Up Time: 3-5 minutes (no flow). Audible Alarm: 75 dB between 3-10 feet. Weight: 3 oz (85 g). Mounting Hardware: (2) #6 x 1" sheet metal screws, (2) 6-32 x 7/8" machine
screws.
POPULAR MODELS
Model Number
Description
660
Air Velocity Monitor, Includes 3' Flexible Tubing,
Pre-fittings and 90º Elbow, 120 VAC to 24 VAC
Power Transformer